ADT760-MA Automatic Handheld Pressure Calibrator

ADT760-MA Automatic Handheld Pressure Calibrator - Prisma Request a quote

Référence : ADT760-MA

Relative & absolut measures Vacuum to 20 Bar

• Fully automatic calibrator with built-in pump and controller
• Switchable internal pressure modules for expandable ranges
• Accuracy (1 year) of 0.02%FS
• External pressure modules available (measure only)
• Less than 4 lbs (1.8 kg) for handheld operation
• Source pressure, measure pressure and electrical
• 4 channels
• Optional HART communications
• Optional data logging and task documenting
• USB, Wi-Fi, and Bluetooth communications

 

 

A portable automated pressure calibrator in the palm of your hand-this could be our most exciting product yet! The Additel 760 series Automatic Handheld Pressure Calibrator takes portable pressure calibration to new levels. Weighing less than 4 lbs (1.8 kg), the ADT760's innovative design contains a built-in pump, precision pressure sensor, internal controller and a large touch-screen color display. To generate pressure, simply key in the desired pressure and the Additel 760 will do the rest. Each unit has four channels: one internal pressure channel for source and measure pressures, two external pressure measurement channels, and one electronic measure and source channel. This series of calibrator has three standard models with the option of adding HART communications, documentation and data logging.

ADT760-MA
The 760-MA generates and controls pressure from vacuum pressures up to 300 psig (20 bar.g) with an accuracy of 0.02%FS. Equipped with a built-in barometric reference, each unit can switch between gauge and absolute pressure types. A variety of internal sensors are available which offer lower pressure ranges for improved performance.

Documenting Process Functionality
Each model of the Additel 760 series has an option incorporating documentation and communication functions turning your 760 into a multifunction documenting process calibrator. This feature provides HART communication, task documentation and data logging.

 

  • Features
  • Specifications
  • Applications
  • Datasheet

• Chanels

Four total: one electrical, one internal pressure, two external pressure (measure only)


• Enclosure IP Rating
IP54 water and dust proof

• Battery
Rechargeable Li-Ion battery, typically 10 hours of operation, recharges in less than 4 hours

• Display
Color 800 x 480 TFT 5-inch touch screen

• Communications
USB, WiFi, Bluetooth

• Weight
<4 lbs (<1.8 kg)

• Size
9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm)

• Certification
NIST-traceable certificate with data included

• HART Communications
Optional (ADT760-X-DL model)

• Data Logging
Optional (ADT760-X-DL model), up to 1,000,000 readings (date and time stamped)

• Task Documentation
Optional (ADT760-X-DL model) up to 250 tasks

• Automation Functions
Switch test, auto step, leak test

• Misuse Protection
Up to 30 V on any two sockets

• Multi Lingual Interface
English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian

• Pump Life

>500,000 cycles

• Power

Rechargeable Li-Ion battery, external power: 110/220 V power adapter 10 V

• Environment Specifications

Operation: 32°F to 122°F (0°C to 50°C), 0-90% RH, less than 3,000 meters
Compensated temperature: 32°F to 122°F (0°C to 50°C)
Storage temperature: -4°F to 158°F (-20°C to 70°C)

• Vibration and Shock
Vibration: 4g (20 to 2,000 Hz)
Shock: 8g, 1 meter drop test

• Compliance
CE

• Warranty
1 year

 

• Max Pressure Range

-12.5 to 300 psig ( -0.86 to 20 bar.g )
• Accuracy
0.02%FS[1][3]
• Stability
<0.005%FS[2]
• Pressure Type
Gauge, Absolute
• Over Range Indication
120%
• Resolution
6 digits
• Measurement Units
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C,
inH2O@20°C, kg/m2, mtorr, torr, lb/ft2, tsi, custom
• Barometric Accuracy
60 Pa[4]
• Connection
G 1/8 BSPF
• Pressure Output Rate
<90 Seconds ( 300 psi / 5 ml )

[1] FS specification applies to the span of the module range.
[2] Stability based on FS of the internal pressure module. Internal module is switchable.
[3] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode.
[4] 60 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long-term stability (<30 Pa per year). Barometer range of 60 to 110 kPa.

 

Optional Internal Pressure Module ADT155 for ADT-760-MAModule de pression ADT155

Improve accuracy of control and measurement


Référence module Description
ADT155-CP10-760 ±10 psi, ±0.02%FS
ADT155-CP15-760 -12.5 to 15 psi, ±0.02%FS
ADT155-CP30-760 -12.5 to 30 psi, ±0.02%FS
ADT155-CP35-760 -12.5 to 35 psi, ±0.02%FS
ADT155-CP50-760 -12.5 to 50 psi, ±0.02%FS
ADT155-CP100-760 -12.5 to 100 psi, ±0.02%FS
ADT155-CP150-760 -12.5 to 150 psi, ±0.02%FS
ADT155-CP200-760 -12.5 to 200 psi, ±0.02%FS
ADT155-CP300-760 -12.5 to 300 psi, ±0.02%FS

 

 

calibrateur de pression portable pour laboratoire

• Differential gauge calibration

• Low pressure calibration

• Differential transmitter calibration

• Differential switch calibration

• HART-smart transmitter calibration ( -DL version only )

• Pressure monitoring and data logging ( -DL version only )

 

Télécharger la fiche technique  (PDF ADT-760)